Hello, Linda
@
VA Wilkes-Barre, PA
|
My Account
|
Support
|
Logout
Manage:
Account Information
First Name
*
Last Name
*
Email / Username
*
Password
*
Preferences
Group Settings
Group name
Please enter a group name.
Assignments Due Date
RadDatePicker
RadDatePicker
Open the calendar popup.
Calendar
Title and navigation
Title and navigation
<<
<
August 2017
>
<<
August 2017
S
M
T
W
T
F
S
31
30
31
1
2
3
4
5
32
6
7
8
9
10
11
12
33
13
14
15
16
17
18
19
34
20
21
22
23
24
25
26
35
27
28
29
30
31
1
2
36
3
4
5
6
7
8
9
Allow users to retake competency tests
Allow users to retake training tests
Allow users to retake document tests
Please select at least one program for your group.
Program
Administrators
Competency Test
i-STAT - ACT
Linda Stubb
Score:
80%
Request Retake
Test Date:
8/28/2017 12:00:22 PM
1
True or False: Cartridges removed from refrigeration can be stored at room temperature for up to 2 weeks.
True
False
Answer required
Flip
Explanation
True. The Kaolin ACT cartridge must be stored at 2-8 °C. Cartridges removed from refrigeration can be stored at room temperature for up to 2 weeks.
2
True or False: Recommended collection options for the i-STAT ACT test include plastic syringes or plastic evacuated tubes each without anticoagulant, clot activators, or serum separators.
True
False
Answer required
Flip
Explanation
True. The recommended collection options for the i-STAT ACT test include plastic syringes or plastic evacuated tubes each without anticoagulant, clot activators, or serum separators. Other collection options are not recommended. Any device used to transfer the specimen to a test cartridge must also be plastic.
3
After collection, when should a specimen be tested?
After 3 minutes of vigorous mixing
Immediately
Within 30 minutes (with anticoagulant)
Answer required
Flip
Explanation
Samples used for the measurement of ACT should be tested immediately after collection.
4
True or False: When drawing blood from an in-dwelling line, the line should be flushed with 5 mL of saline and the first 5 mL of blood (or 6 dead space volumes) should be discarded before drawing the patient specimen for testing.
True
False
Answer required
Explanation
When drawing blood from an in-dwelling line, the line should be flushed with 5 mL of saline and the first 5 mL of blood (or 6 dead space volumes) should be discarded before drawing the patient specimen for testing.
5
When should ACT controls be used to validate cartridges?
Only when a recurrent problem arises
With every new shipment of cartridges
Every six months
Daily
Answer required
Flip
Explanation
All new shipments of cartridges must be validated with ACT controls prior to use with patient specimens.
Request Retake
Your request will be sent to the Administrator selected below